4.13 Time Domain Graph Display
4-195
(4) Marker data setting
Select or clear the Data Edit check box to specify whether to edit the
data of the corresponding marker. When it is cleared, marker data
edit is disabled and the marker data of the read waveform pattern is
used as is. When it is selected, marker data edit is enabled, and the
marker data with the values set in the Start Point, Width, and
Period text boxes are added to the new waveform pattern to be
created.
Set the number of samples prior to the first marker signal output
starting point to Start Point, the number of samples to output
marker signals to Width, and the number of samples between the
marker signal output starting points to Period.
Start Point Width
Waveform pattern
Marker (RF Gate)
Width
Width
Period
Period
Period
Figure 4.13.12-2 Setting Parameters of Marker and RF Gate Data
(5) RF Gate setting
Select or clear the Data Edit check box to specify whether to edit the
data of the corresponding marker. When it is cleared, RF Gate data
edit is disabled and the RF Gate data of the read waveform pattern
is used as is. When it is selected, RF Gate data edit is enabled, and
the RF Gate data with the values set in the Start Point, Width, and
Period text boxes are added to the new waveform pattern to be
created.
Set the number of samples prior to the first marker signal output
starting point to Start Point, the number of samples to output
marker signals to Width, and the number of samples between the
marker signal output starting points to Period.