DBS9900 Registers
6-5
82-28993 Revision 01
6.3 Control Register
The CONTROL REGISTER (0x04) is used to generate a SOFT RESET,
•
inhibit SYSFAIL generation
•
to enable A24/A32 memory access.
When the SOFT RESET bit is set, it should be cleared again with the next operation to the
DBS9900. No other operations should be attempted with the SOFT RESET bit SET, or a bus
error will result.
Table 3 - CONTROL REGISTER - WRITE (0x04)
BIT
Name
Set Function
Clear Function
D0
SOFT RESET
Resets DBS9900 registers
Enables access to DBS9900
D1
SYSFAIL INHB
Prevents SYSFAIL assertion
Enables SYSFAIL assertion
D2 –D14
Not Used
Not Used
D15
A24/A32 ENABLE
Enables DBS9900 memory access
Disconnects DBS9900 memory
6.4 Status Register
The STATUS REGISTER (0x04) is used to read back the CONTROL REGISTER bits: READY,
PASSED, and VXI MODID. It is also used to detect that a daughter board is present and its
operational status. The Status Register has four additional bits added to the basic VXI functions.
These are read only. Two bits are used to indicate that a module is present (i.e. installed) on the
DBS9900. One bit is assigned for each possible module. If the module is present its
corresponding bit will be SET. This defines up to two modules per VXI Motherboard. Two
additional bits are used as DONE bits, one for each module. The Done bits will be duplicated in
the module STATUS registers for flexibility.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com