Maintenance and Troubleshooting
|
6-5
Memory is Corrupted
This error message indicates that the control unit’s internal EEPROM memory has
been corrupted, or a significantly new version of software has been installed in the
control unit. Under these conditions, the software will reprogram the EEPROM
with the factory default values. You can acknowledge this error message by setting
or reading any user programmable value, and this will clear the message from the
display. This message will not clear by removing and restoring power to the unit.
If you receive this message, you should reset your control unit to the values required
for your application, including Setup key parameters, Calibrate key parameters, etc...
Span Gas Range Error
This error message occurs if the calibration span gas does not read within the soft-
ware’s allowable limits. Once this error occurs, calibration of the system is aborted.
This message could indicate one of the following problems:
Calibration gas setup problem
Process pressure incorrectly entered
Failed cell
Cell wiring reversed (Cell + and Cell - wires)
Leak in plumbing
Plugged plumbing
Содержание CEM/Humox
Страница 20: ...2 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 38: ...3 18 Thermox CEM Humox Analyzer Figure 3 16 Series 2000 alarm connections example ...
Страница 40: ...3 20 Thermox CEM Humox Analyzer Figure 3 17 Series 2000 2 wire RS 485 communication connection ...
Страница 42: ...3 22 Thermox CEM Humox Analyzer Figure 3 18 Series 2000 4 wire RS 485 communication connections ...
Страница 44: ...3 24 Thermox CEM Humox Analyzer ...
Страница 76: ...4 32 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 105: ...Service and Parts 7 3 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 107: ...Service and Parts 7 5 Figure 7 1 Cell replacement ...
Страница 130: ...B 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 134: ...C 4 Thermox CEM Humox Analyzer Figure C 1 Modulation of external power supply OPTIONAL ...
Страница 136: ...D 2 Thermox CEM Humox Analyzer This page intentionally left blank ...