A-8 | Thermox CEM/Humox Analyzer
Variable Table
The following pages provide a summary of the variable locations, names, and
descriptions available. If you need access to more advanced variables, please call
AMETEK.
Location
(Hex/Decimal)
Variable Name
Description
Access
00h/000d
Flags3
Various flags (1=true, 0=false)
bit 0 - sensor is at operating temperature.
bit 1 - over/under flag for tuning the temperature.
bit 2 - inject cal gas active.
bit 3 - auto calibrate is occurring.
bit 4 - auto verify is occurring.
bit 5 - in a menu
bit 6 - No limit msg
bit 7 - calibrate or verify is not permitted
Read only
01h/001d
Inf_mess_flag
+1
+2
+3
System and error message flags (1=true, 0 = false)
bit 0 - temperature rise failure.
bit 1 - cell is over temperature.
bit 2 - zero gas range error, true when (ideal span mv
- ideal zero mv) - (measured span mv - measured zero
mv) > 5mv.
bit 3 - span gas range error, true when (ideal span
gas mv) - (measured zero gas mv) > 30 mv
bit 4 - primary calibration being performed
bit 5 - memory is corrupted, meaning the EEPROM has
been erased and is using the default values.
bit 6 - excessive cal error, true when the primary
calibration cannot adjust the temperature of the cell
enough to make the cell read the correct mv. The up
-
per actual temperature for the insitu is 800C, for the
3 in 1, 925C, and for all others, 850C.
bit 7 - calibration required, true after a primary cal,
or if the EEPROM has been erased and the unit is using
default values.
bit 8 - thermocouple failure
bit 9 - thermocouple compensation failure
bit 10 - calibration or verify aborted.
bit 11 - calibration aborted.
bit 12 - system verifying calibration, will be true
during the verify and any recovery time that is speci
-
fied.
bit 13 - system calibrating, will be true during the
calibration and during any recovery time that is speci
-
fied.
bit 14 - auto verify pending, an auto verify is going
to occur at the set time
bit 15 - auto calibration pending, an auto calibration
is going to occur at the set time
bit 16 - verify failure, the verify gas has measured
outside the range set by the user.
bit 17 - Future use
bit 18 - Future use
bit 19 - Future use
bit 20 - Future use
bit 21 -
bit 22 - Future use
bit 23 - Future use
bit 24 - Future use
bit 25 - Future use
bit 26 - Future use
bit 27 - Low oxygen
bit 28 - Power down detected
bit 29 - Warming up
bit 30 – Reserved
bit 31 – Reserved
Read only
Содержание CEM/Humox
Страница 20: ...2 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 38: ...3 18 Thermox CEM Humox Analyzer Figure 3 16 Series 2000 alarm connections example ...
Страница 40: ...3 20 Thermox CEM Humox Analyzer Figure 3 17 Series 2000 2 wire RS 485 communication connection ...
Страница 42: ...3 22 Thermox CEM Humox Analyzer Figure 3 18 Series 2000 4 wire RS 485 communication connections ...
Страница 44: ...3 24 Thermox CEM Humox Analyzer ...
Страница 76: ...4 32 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 105: ...Service and Parts 7 3 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 107: ...Service and Parts 7 5 Figure 7 1 Cell replacement ...
Страница 130: ...B 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 134: ...C 4 Thermox CEM Humox Analyzer Figure C 1 Modulation of external power supply OPTIONAL ...
Страница 136: ...D 2 Thermox CEM Humox Analyzer This page intentionally left blank ...