Installation
|
3-3
Sample Gas Requirements/Connection
Figure 3-1 shows the inlet and outlet ports on the CEM/O
2
sensor. The sample gas
should be clean and dry. If the sample contains any condensing moisture, the entire
sample line should be heat traced.
The sensor may operate at a steady internal pressure up to 20 WC (5 kPa). Outlet
restrictions that raise the pressure beyond this level may cause false or unstable
readings. The sample gas flow rate must be between 2 and 20 SCFH (0.94 to 9.4 L/
min). The maximum sample temperature is 204°C (400 °F).
Figure 3-1.
Sensor mounting.