Controller/User Interface
|
4-27
You are then prompted to enter the alarm 6 set point value:
Alarm #6 = xx
New Value?
5. Enter the alarm 6 set point value.
6. Press the Cancel key or the Enter key to then exit this menu option.
Alarm Configure
This menu option allows you to define whether alarms should trigger based on a
high alarm or a low alarm condition, and also to select what reading to track. See
the “Alarm Set Points” section for help on how to define alarm set point values.
Whether Alarm 3 is set high or low only applies if it is set to track sen-
sor readings. If Alarm 3 is set to trigger at the start of a calibration or
verification, this option does not apply.
1. Select Alarm Configure from the Alarm menu.
2. Define your alarms as high or low alarms. (See the “Alarm Set Point” section
in this chapter for information on setting alarm set points.)
3. Press the Cancel key to then exit this menu option.
NOTE
Содержание CEM/Humox
Страница 20: ...2 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 38: ...3 18 Thermox CEM Humox Analyzer Figure 3 16 Series 2000 alarm connections example ...
Страница 40: ...3 20 Thermox CEM Humox Analyzer Figure 3 17 Series 2000 2 wire RS 485 communication connection ...
Страница 42: ...3 22 Thermox CEM Humox Analyzer Figure 3 18 Series 2000 4 wire RS 485 communication connections ...
Страница 44: ...3 24 Thermox CEM Humox Analyzer ...
Страница 76: ...4 32 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 105: ...Service and Parts 7 3 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 107: ...Service and Parts 7 5 Figure 7 1 Cell replacement ...
Страница 130: ...B 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 134: ...C 4 Thermox CEM Humox Analyzer Figure C 1 Modulation of external power supply OPTIONAL ...
Страница 136: ...D 2 Thermox CEM Humox Analyzer This page intentionally left blank ...