Installation
|
3-17
Alarm contact connections
This section describes how to make wiring connections for any alarm devices you
wish to connect to the control unit. Information on how to set up alarms in software
is described in the Alarms Key section.
Alarm connections
The Series 2000 control unit provides six sets of standard alarm contacts, which are
labeled as follows (see Figure 3-16):
ALARM 1A/1B through ALARM 6A and 6B
The
Watchdog Timer Alarm
connections are labeled as follows on the control unit:
ALARM 1A
ALARM 1B
See the “Watchdog ”Relay” in the Alarm Key section for complete information on
the watchdog timer alarm. The watchdog alarm contacts are normally closed during
system operations (deactivated). A watchdog event triggers these alarm contacts to
an open condition. If you have a watchdog event, and the system returns to normal
operations, then the watchdog alarm will reset. The system will retain in memory
the last time the watchdog condition occurred (see the “Exception Log” section in
the Alarm Key section for help on how to view this information).
The
Service Alarm
connections are labeled as follows:
ALARM 2A
ALARM 2B
See the “Service Alarm” in the Alarm Key section for complete information on
what events trigger the transfer of the service alarm. The service alarm contacts
are normally closed during system operations (deactivated). A service alarm event
triggers these alarm contacts to an open condition.
Sensor Alarm
connections are configurable in software to activate under a variety
of user defined conditions, and are labeled as follows:
ALARM 3A ALARM 4A ALARM 5A ALARM 6A
ALARM 3B ALARM 4B ALARM 5B ALARM 6B
See the Alarm Key section for complete information on all alarms, including what
they can track and how they operate.
Содержание CEM/Humox
Страница 20: ...2 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 38: ...3 18 Thermox CEM Humox Analyzer Figure 3 16 Series 2000 alarm connections example ...
Страница 40: ...3 20 Thermox CEM Humox Analyzer Figure 3 17 Series 2000 2 wire RS 485 communication connection ...
Страница 42: ...3 22 Thermox CEM Humox Analyzer Figure 3 18 Series 2000 4 wire RS 485 communication connections ...
Страница 44: ...3 24 Thermox CEM Humox Analyzer ...
Страница 76: ...4 32 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 105: ...Service and Parts 7 3 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 107: ...Service and Parts 7 5 Figure 7 1 Cell replacement ...
Страница 130: ...B 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 134: ...C 4 Thermox CEM Humox Analyzer Figure C 1 Modulation of external power supply OPTIONAL ...
Страница 136: ...D 2 Thermox CEM Humox Analyzer This page intentionally left blank ...