Serial Communications
|
A-9
02h/002d
Config_flags
System configuration flags (1=true, 0=false)
bit 0 - relays energize on alarm. If not, they deen
-
ergize.
bit 1 - auto calibration time is set, an auto calibra
-
tion will happen at the set time if the auto calibra
-
tion is enabled
bit 2 - auto calibration timer is enabled, since this
is enabled, an auto calibration will occur at the auto
calibration time.
bit 3 - real time clock is set.
bit 4 - Future use
bit 5 - auto verify time is set, an auto verify will
happen at the set time if the auto verify is enabled.
bit 6 - auto verify timer is enabled, if this is en
-
abled an auto verify will occur at the auto verify
time.
bit 7 - auto calibration cycle is programmed in hours,
0 if in days
bit 8 - auto verify cycle time is programmed in hours,
0 if in days
bit 9 - auto calibrate option is installed
bit 10 - Reserved
bit 11 - Future use
bit 12 - Future use
bit 13 - Future use
bit 14 - Future use
bit 15 - Future use
Read and Write
03h/003d
Iout_flags
Analog outputs flags (1=true, 0=false)
bit 0 - analog output 1 tracks during calibration, 0
is hold analog output 1 during calibration
bit 1 - track analog output 1 during verify, 0 is hold
analog output during verify.
bit 2 - analog output 1 set for 0 to 20 ma, 0 is 4 to
20 ma
bit 3 - Future use
bit 4 - analog output 2 tracks during calibration, 0
is hold analog output 2 during calibration
bit 5 - analog output 2 tracks during verify, 0 is hold
analog output during verify.
bit 6 - analog output 2 set for 0 to 20 ma, 0 is 4 to
20 ma.
bit 7 - Future use
bit 8 - analog output 3 tracks during calibration, 0
is hold analog output 3 during calibration
bit 9 - analog output 3 tracks during verify, 0 is hold
analog output 3 during verify.
bit 10 - analog output 3 set for 0 to 20 ma, 0 is 4 to
20 ma
bit 11 - Future use
bit 12 - analog output 4 tracks during calibration, 0
is hold analog output 4 during calibration.
bit 13 - analog output 4 tracks during verify, 0 is
hold analog output 4 during verify.
bit 14 - analog output 4 set for 0 to 20 ma, 0 is 4 to
20 ma
bit 15 - Future use
Read and Write
04h/004d
Reserved
N/A
05h/005d
Dsp1_flg
What is displayed on line 1 of the display
0 - dry oxygen
2 - dry cell temperature
4 - dry thermocouple mv
6 - dry cell mv
8 - wet oxygen
10 - wet cell temperature
12 - wet thermocouple mv
14 - wet cell mv
16 - moisture
Read and Write
06h/006d
Dsp2_flg
What is displayed on line 2 of the display
Same as Line1_flg
Read and Write
07h/007d
Dsp3_flg
What is displayed on line 3 of the display
Same as Line1_flg
Read and Write
08h/008d
Dry_Oxygen
The Dry oxygen reading in percent
Read only
09h/009d
Reserved
N/A
0Ah/010d
Reserved
N/A
0Bh/011d
Dry_Cell_temp
The current Dry cell temperature in degrees C
Read only
0Ch/012d
Dry_Cell_mv
The Dry cell millivolts
Read only
0Dh/013d
Dry_Tc_mv
The Dry thermocouple millivolts
Read only
Содержание CEM/Humox
Страница 20: ...2 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 38: ...3 18 Thermox CEM Humox Analyzer Figure 3 16 Series 2000 alarm connections example ...
Страница 40: ...3 20 Thermox CEM Humox Analyzer Figure 3 17 Series 2000 2 wire RS 485 communication connection ...
Страница 42: ...3 22 Thermox CEM Humox Analyzer Figure 3 18 Series 2000 4 wire RS 485 communication connections ...
Страница 44: ...3 24 Thermox CEM Humox Analyzer ...
Страница 76: ...4 32 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 105: ...Service and Parts 7 3 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 107: ...Service and Parts 7 5 Figure 7 1 Cell replacement ...
Страница 130: ...B 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 134: ...C 4 Thermox CEM Humox Analyzer Figure C 1 Modulation of external power supply OPTIONAL ...
Страница 136: ...D 2 Thermox CEM Humox Analyzer This page intentionally left blank ...