3-16 | Thermox CEM/Humox Analyzer
Current output connections
Current output connections are labeled as follows on the wiring card of the Series
2000 control unit (see Figure 3-15):
IOUT1+ and -
IOUT3+ and -
IOUT4+ and -
These current outputs are referenced as current outputs 1, 3, and 4 in software (see
the Analog Range Key section for help on defining current output settings in soft
-
ware). Be sure to observe polarity when connecting current output devices to these
terminals. Analog output 2 is reserved for furnace control.
Current outputs 1, 3 and 4 are capable of driving up to 1200 ohm loads. These cur
-
rent outputs are isolated from the control unit, but are not isolated from each other.
If you intend to power current outputs from an external power supply,
see the “External Powering and Modulation of External Power Using
Current Outputs” appendix.
NOTE
Figure 3-15.
Current output
connections example.
Содержание CEM/Humox
Страница 20: ...2 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 38: ...3 18 Thermox CEM Humox Analyzer Figure 3 16 Series 2000 alarm connections example ...
Страница 40: ...3 20 Thermox CEM Humox Analyzer Figure 3 17 Series 2000 2 wire RS 485 communication connection ...
Страница 42: ...3 22 Thermox CEM Humox Analyzer Figure 3 18 Series 2000 4 wire RS 485 communication connections ...
Страница 44: ...3 24 Thermox CEM Humox Analyzer ...
Страница 76: ...4 32 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 105: ...Service and Parts 7 3 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 107: ...Service and Parts 7 5 Figure 7 1 Cell replacement ...
Страница 130: ...B 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 134: ...C 4 Thermox CEM Humox Analyzer Figure C 1 Modulation of external power supply OPTIONAL ...
Страница 136: ...D 2 Thermox CEM Humox Analyzer This page intentionally left blank ...