6-6 | Thermox CEM/Humox Analyzer
Temp Rise Failure (Wet or Dry Cell)
This error message indicates that the wet or dry sensor has failed to increase in
temperature a minimum of 10°C in a 60 second time frame during start-up. Once
the system reaches the operating set point, this message will only occur if the sen-
sor falls below the set point by 15°C and fails to recover within 60 seconds. The
problems that can cause this error are as follows:
AC line power problem at the sensor
Open furnace
Over-temperature protection circuit trip on the sensor board
Sensor board failure
Loss of furnace drive signal
Interconnecting wiring problem
T/C (Thermocouple) Circuit Failure
This message appears when the control unit recognizes a sudden drop in temperature
(greater than 100°C). The thermocouple circuit failure message can be the result of
one of the following problems:
Shorted thermocouple
Faulty interconnecting wiring
Control unit display module failure
Sensor board failure
If this message is soon followed by a “Thermocouple Failure” message, see that
message for troubleshooting instructions.
Содержание CEM/Humox
Страница 20: ...2 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 38: ...3 18 Thermox CEM Humox Analyzer Figure 3 16 Series 2000 alarm connections example ...
Страница 40: ...3 20 Thermox CEM Humox Analyzer Figure 3 17 Series 2000 2 wire RS 485 communication connection ...
Страница 42: ...3 22 Thermox CEM Humox Analyzer Figure 3 18 Series 2000 4 wire RS 485 communication connections ...
Страница 44: ...3 24 Thermox CEM Humox Analyzer ...
Страница 76: ...4 32 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 105: ...Service and Parts 7 3 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 107: ...Service and Parts 7 5 Figure 7 1 Cell replacement ...
Страница 130: ...B 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 134: ...C 4 Thermox CEM Humox Analyzer Figure C 1 Modulation of external power supply OPTIONAL ...
Страница 136: ...D 2 Thermox CEM Humox Analyzer This page intentionally left blank ...