Controller/User Interface
|
4-9
Dry Process Pressure
This menu option allows you to define your dry process pressure. It also allows you
to define whether your process is operating under a positive or negative pressure.
Setting this value corrects dry oxygen measurements for the defined pressure.
1. Select Dry Process Press from the Setup key menu.
2. Select Positive Pressure or Negative Pressure.
You are then prompted to enter your system process pressure:
Pressure = xx.x PSIG
New Value?
3. Enter your process pressure (up to one decimal point is allowed - for example,
1.5 PSIG), then press the Enter key. If you select a positive pressure, a plus sign
(+) will appear before your value. If you select a negative pressure, a negative
sign (-) will appear before your value.
The process pressure you entered will now appear on the third display line.
4. If the correct pressure has been entered into the system, press the Enter key
without entering a new value to exit this menu option. If the pressure was incor-
rectly entered, you can reenter it at this point.
Содержание CEM/Humox
Страница 20: ...2 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 38: ...3 18 Thermox CEM Humox Analyzer Figure 3 16 Series 2000 alarm connections example ...
Страница 40: ...3 20 Thermox CEM Humox Analyzer Figure 3 17 Series 2000 2 wire RS 485 communication connection ...
Страница 42: ...3 22 Thermox CEM Humox Analyzer Figure 3 18 Series 2000 4 wire RS 485 communication connections ...
Страница 44: ...3 24 Thermox CEM Humox Analyzer ...
Страница 76: ...4 32 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 105: ...Service and Parts 7 3 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 107: ...Service and Parts 7 5 Figure 7 1 Cell replacement ...
Страница 130: ...B 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 134: ...C 4 Thermox CEM Humox Analyzer Figure C 1 Modulation of external power supply OPTIONAL ...
Страница 136: ...D 2 Thermox CEM Humox Analyzer This page intentionally left blank ...