C is calculated from the following equation:
C = % residual moisture = (Psat/14.696) * 100%
Psat (saturated pressure) determined from Saturated Water - Temperature table
knowing the vortex cooler temperature. A regression was performed from the look-
up table data for temperatures from 0°C to 32.2°C (32°F to 90° F). The resulting
function is sued in the Humox software to calculate the residual moisture.
C = % residual moisture = 0.63e(.064 * Cooler Temperature in C)
The wet and dry O
2
(% by Volume) are calculated using the following Nernst equa
-
tions:
% O
2
wet = 20.9% O
2
wet / 10 (wet cell mV/AT)
% O
2
dry = 20.9% O
2
dry / 10 (dry cell mV/AT)
AT - Cell logarithmic slope (ideally 48 mV per decade for T = 695°C)
A - Constant (.0496 mV/decade/°K)
T - Temperature (°K)(°C + 273)
Содержание CEM/Humox
Страница 20: ...2 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 38: ...3 18 Thermox CEM Humox Analyzer Figure 3 16 Series 2000 alarm connections example ...
Страница 40: ...3 20 Thermox CEM Humox Analyzer Figure 3 17 Series 2000 2 wire RS 485 communication connection ...
Страница 42: ...3 22 Thermox CEM Humox Analyzer Figure 3 18 Series 2000 4 wire RS 485 communication connections ...
Страница 44: ...3 24 Thermox CEM Humox Analyzer ...
Страница 76: ...4 32 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 105: ...Service and Parts 7 3 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 107: ...Service and Parts 7 5 Figure 7 1 Cell replacement ...
Страница 130: ...B 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 134: ...C 4 Thermox CEM Humox Analyzer Figure C 1 Modulation of external power supply OPTIONAL ...
Страница 136: ...D 2 Thermox CEM Humox Analyzer This page intentionally left blank ...