3-12 | Thermox CEM/Humox Analyzer
Transient and RFI interference
This section describes transient and RFI interference precautions:
• Although there are transient and noise protectors on all control unit I/O connec-
tions (communications, current outputs, sensor, etc ...), this protection is intended
to act as a last line of defense against unwanted transient and RFI interference.
Proper installation practices to prevent the introduction of transients and noise
into the system must be followed. Inductive loads connected to the control unit
must have transient suppressers installed at the inductive loads. Be sure to place
the transient suppresser as close to the load as possible. Examples of transient
suppressers include MOVs, TRANSORBs, and RC snubbers.
• AC mains supply wiring should not be run in the same conduit with mains sup-
ply wiring that feeds heavy inductive loads.
• Avoid running signal wiring in the same cable or conduit with wires that power
inductive loads unless all the cables within the conduit are shielded, the induc-
tive loads are small, and transient suppressers are used at the loads.
• Do not run signal lines in the same cable or conduit with high voltage lines.
• For optimum noise protection, control unit mains supply wiring should be
connected to a circuit separate from any circuit that could introduce transients
into the system. As an example, do not run motors, blowers, or air condition-
ers using the same mains supply circuit or conduit as the control unit’s mains
supply circuit or conduit.
Figure 3-11.
Direct shield
grounding methods.
SHIELD RING METHOD
GROUND STUD METHOD
Содержание CEM/Humox
Страница 20: ...2 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 38: ...3 18 Thermox CEM Humox Analyzer Figure 3 16 Series 2000 alarm connections example ...
Страница 40: ...3 20 Thermox CEM Humox Analyzer Figure 3 17 Series 2000 2 wire RS 485 communication connection ...
Страница 42: ...3 22 Thermox CEM Humox Analyzer Figure 3 18 Series 2000 4 wire RS 485 communication connections ...
Страница 44: ...3 24 Thermox CEM Humox Analyzer ...
Страница 76: ...4 32 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 105: ...Service and Parts 7 3 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 107: ...Service and Parts 7 5 Figure 7 1 Cell replacement ...
Страница 130: ...B 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 134: ...C 4 Thermox CEM Humox Analyzer Figure C 1 Modulation of external power supply OPTIONAL ...
Страница 136: ...D 2 Thermox CEM Humox Analyzer This page intentionally left blank ...