4-24 | Thermox CEM/Humox Analyzer
Alarm Key
Alarms 1 and 2 are hard wired for the watchdog timer and service
alarms and are therefore not selectable in software.
Sensor alarms are all disabled during a primary or regular calibra-
tion. In addition, alarms are disabled during start-up until the sensor
reaches its proper operating temperature.
See the “Wiring” section in the Installation chapter for help on
alarm connections and how alarms are labeled on the control unit
wiring card.
The Alarm key allows you to do the following:
Define whether alarm 3 activates based on oxygen levels, the start of a calibra
-
tion, or the start of a verification (Select Function). Other alarms activate only
on sensor readings.
Define whether alarms should be set high or low (Alarm Configure).
Define whether alarm relays should energize or deenergize on alarm (Relay
Configure). The selection you make using the Relay Configure menu option
applies to all alarms except the watchdog and service alarms, which always
deenergize on alarm.
Figure 4-6.
Alarm Key functions.
NOTE
Alarm Set Points
Alarm Configure
Relay Configure
Exception Log
Select Function
Alarm Key
Содержание CEM/Humox
Страница 20: ...2 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 38: ...3 18 Thermox CEM Humox Analyzer Figure 3 16 Series 2000 alarm connections example ...
Страница 40: ...3 20 Thermox CEM Humox Analyzer Figure 3 17 Series 2000 2 wire RS 485 communication connection ...
Страница 42: ...3 22 Thermox CEM Humox Analyzer Figure 3 18 Series 2000 4 wire RS 485 communication connections ...
Страница 44: ...3 24 Thermox CEM Humox Analyzer ...
Страница 76: ...4 32 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 105: ...Service and Parts 7 3 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 107: ...Service and Parts 7 5 Figure 7 1 Cell replacement ...
Страница 130: ...B 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 134: ...C 4 Thermox CEM Humox Analyzer Figure C 1 Modulation of external power supply OPTIONAL ...
Страница 136: ...D 2 Thermox CEM Humox Analyzer This page intentionally left blank ...