A-12 | Thermox CEM/Humox Analyzer
63h/099d
Option_flags
Option flags (1=true, 0=false)
bit 0 - reserved
bit 1 - dual current output option installed
bit 2 - auto cal option installed
bit 3 - three-in-one option installed
bit 4 - dual current & furnace card
bit 5 - Future use
bit 6 - Future use
bit 7 - invalid option module
Read only
64h/100d
Dry_Tspan_mv
Dry Cell mv value measured with the span gas flowing
during last calibration.
Read only
65h/101d
Dry_Tzero_mv
Dry Cell mv value measured with the zero gas flowing
during last calibration.
Read and Write
66h/102d
Reserved
N/A
67h/103d
Dry_M_store
The Dry furnace addition value
Read only
68h/104d
Dry_Cal_temp
Dry calibration temperature
Read only
69h/105d
Wet_Oxygen
The Wet oxygen reading in percent
Read only
6Ah/106d
Wet_Cell_temp
The current Wet cell temperature in degrees C
Read only
6Bh/107d
Wet_Cell_mv
The Wet cell millivolts
Read only
6Ch/108d
Wet_Tc_mv
The Wet thermocouple millivolts
Read only
6Dh/109d
Wet_Cal_cgas1
Wet Oxygen span gas setpoint for last calibration
Read only
6Eh/110d
Wet_Cal_span
Wet Oxygen span gas reading for last calibration
Read only
6Fh/111d
Wet_Ver_cgas1
Wet Oxygen span gas setpoint for last verify
Read only
70h/112d
Wet_Ver_span
Wet Oxygen span gas reading from last verify
Read only
71h/113d
Wet_O2_span
Wet Oxygen span gas millivolts during last calibration
Read only
72h/114d
Wet_Cal_cgas2
Wet Oxygen zero gas setpoint value for last calibration
Read only
73h/115d
Wet_Cal_zero
Wet Oxygen zero gas reading from last calibration
Read only
74h/116d
Wet_Ver_cgas2
Wet Oxygen setpoint value used during last verify
Read only
75h/117d
Wet_Ver_zero
Wet Oxygen zero gas read during last verify
Read only
76h/118d
Wet_O2_zero
Wet Zero gas millivolts read during last calibration
Read only
77h/119d
Reserved
N/A
78h/120d
Reserved
N/A
79h/121d
Wet_AT
Wet Cell slope
Read only
7Ah/122d
Wet_AT_ratio
Wet Cell slope used during last calibration
Read only
7Bh/123d
Wet_Tc_ratio
Cell/thermocouple compensation multiplier
Read only
7Ch/124d
Wet_Cal_variable
The wet calculated percent value that 0 mv would cor
-
respond to with the cell that is currently being used.
This value can vary between different cells. It is
calculated during calibration and used in the wet oxy
-
gen calculations.
Read only
7Dh/125d
Wet_Tspan_mv
Wet Cell mv value measured with the span gas flowing
during last calibration.
Read only
7Eh/126d
Wet_Tzero_mv
Wet Cell mv value measured with the zero gas flowing
during last calibration.
Read only
7Fh/127d
Wet_M_store
The Wet furnace addition value
Read only
80h/128d
Wet_Cal_temp
Wet calibration temperature
Read only
81h/129d
Moisture
% Moisture value
Read only
82h/130d
Cooler Temperature
The current temperature of the vortex cooler
Read only
83h/131d
Wet_Press
Wet Process pressure, entered by the user
Read and Write
84h/132d
Resid_Mst
The Residue moist value
Read only
85h/133d
Alm4_flg
Configuration for alarm 4, high or low. If a bit is 1,
that is the condition that the indicated alarm will
trip on. Same as Alm3_flg
Read and Write
86h/134d
Alm5_flg
Configuration for alarm 5, high or low. If a bit is 1,
that is the condition that the indicated alarm will
trip on. Same as Alm3_flg
Read and Write
87h/135d
Alm6_flg
Configuration for alarm 6, high or low. If a bit is 1,
that is the condition that the indicated alarm will
trip on. Same as Alm3_flg
Read and Write
Содержание CEM/Humox
Страница 20: ...2 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 38: ...3 18 Thermox CEM Humox Analyzer Figure 3 16 Series 2000 alarm connections example ...
Страница 40: ...3 20 Thermox CEM Humox Analyzer Figure 3 17 Series 2000 2 wire RS 485 communication connection ...
Страница 42: ...3 22 Thermox CEM Humox Analyzer Figure 3 18 Series 2000 4 wire RS 485 communication connections ...
Страница 44: ...3 24 Thermox CEM Humox Analyzer ...
Страница 76: ...4 32 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 105: ...Service and Parts 7 3 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 107: ...Service and Parts 7 5 Figure 7 1 Cell replacement ...
Страница 130: ...B 4 Thermox CEM Humox Analyzer This page intentionally left blank ...
Страница 134: ...C 4 Thermox CEM Humox Analyzer Figure C 1 Modulation of external power supply OPTIONAL ...
Страница 136: ...D 2 Thermox CEM Humox Analyzer This page intentionally left blank ...