Compatible SCPI Commands
Supported SCPI Commands
CALCulate<1|2>:LIMit<1…6>:ACPower:ACHannel:ABSolute:STATe
Invokes the limit check for the adjacent channel when ACP measurement is performed. Before the
command, the limit check for the channel/adjacent-channel measurement must be globally switched on
using CALC:LIM:ACP ON.
The result can be queried with CALCulate:LIMit:ACPower:ACHannel:RESult?. A complete
measurement must be performed between switching on the limit check and the result query, since
otherwise no valid results are available.
The numeric suffix in LIMIt is irrelevant for this command.
CALCulate<1|2>:LIMit<1…6>:ACPower:ACHannel:RESult?
Queries the result of the limit check for the upper/lower adjacent channel in the selected measurement
window when ACP measurement is performed.
If the power measurement of the adjacent channel is switched off, the command produces a query error.
The numeric suffix is irrelevant for this command.
The result is returned in the form <result>, <result> where <result> = PASSED | FAILED, and where the
first returned value denotes the lower, the second denotes the upper adjacent channel.
Example
CALC:LIM:ACP:ACH:ABS 30DBM, 30DBM
CALC:LIM:ACP:ACH:ABS?
Notes
FSU, FSP, FSE
Preset
–200dBm, –200dBm
State Saved
Saved in instrument state.
Remote Command
:CALCulate[1]:LIMit[1]|2|3|4|5|6:ACPower:ACHannel:ABSol
ute:STATe ON|OFF|1|0
:CALCulate[1]:LIMit[1]|2|3|4|5|6:ACPower:ACHannel:ABSol
ute:STATe?
Example
CALC:LIM:ACP:ACH:ABS:STAT ON
CALC:LIM:ACP:ACH:ABS:STAT?
Notes
FSU, FSP, FSE
Preset
OFF
State Saved
Saved in instrument state.
Remote Command
:CALCulate[1]:LIMit[1]|2|3|4|5|6:ACPower:ACHannel:RESul
t?
Example
CALC:LIM:ACP:ACH:RES?
Notes
FSU, FSP, FSE