![Agilent Technologies CXA N9000A Скачать руководство пользователя страница 133](http://html.mh-extra.com/html/agilent-technologies/cxa-n9000a/cxa-n9000a_user-manual_2868013133.webp)
132
Compatible SCPI Commands
Supported SCPI Commands
[SENSe<1|2>:]BANDwidth|BWIDth[:RESolution]:TYPE NORMal | FFT | CFILter | RRC
Sets the resolution bandwidth filter type to "normal" analog, FIR in 1, 3, 10 steps, or FFT for bandwidths
less than 100 kHz.
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo 1Hz to 10MHz
Sets the instrument's video bandwidth to between10 Hz and 10 MHz in 1, 3, 10 steps.
In the SCPILC mode, the maximum video bandwidth is 50 MHz and not 10 MHz as in FSx.
Remote Command Notes
FSE, FSP, FSU
Preset
0.02 with BAND:TYPE NORMal or RBW > 30 kHz
0.01 with BAND:TYPE FFT for RBW 30 kHz
State Saved
Saved in instrument state.
Min
0.0001
Max
1
Backwards Compatibility SCPI
:SENSe1:BANDwidth|BWIDth[:RESolution]:RATio
Remote Command
[:SENSe]:BANDwidth|BWIDth[:RESolution]:TYPE
NORMal|FFT|CFILter | RRC
[:SENSe]:BANDwidth|BWIDth[:RESolution]:TYPE?
Example
BAND:TYPE RRC
BAND:TYPE?
Remote Command Notes
FSP, FSU
Preset
NORMal
State Saved
Saved in instrument state.
Range
NORMal | FFT | CFILter | RRC
Backwards Compatibility SCPI
:SENSe1:BANDwidth|BWIDth[:RESolution]:TYPE
Remote Command
[:SENSe]:BANDwidth|BWIDth:VIDeo <freq>
[:SENSe]:BANDwidth|BWIDth:VIDeo?
Example
BAND:VID 1KHZ
BAND:VID?
Remote Command Notes
FSE, FSP, FSU
Couplings
Preset
10 MHz
State Saved
Saved in instrument state.
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html