119
Compatible SCPI Commands
Supported SCPI Commands
The query returns the current value of this setting.
INPut<1|2>:ATTenuation:AUTO
The command switches between coupling of the input attenuation to the reference level (state ON), and
manual input attenuation (state OFF).
The query returns the current state of this setting.
INPut<1|2>:EATT
The command specifies the setting of the electronic input attenuator.
This mode supports a maximum value of 24 dB.
The query returns the current value of this setting.
Remote Command
:INPut[1]:ATTenuation <rel_ampl>
:INPut[1]:ATTenuation?
Example
INP:ATT 40dB
Remote Command Notes
FSE, FSP, FSU
Preset
6dB
State Saved
Saved in instrument state.
Min
0
Max
70
Remote Command
:INPut[1]:ATTenuation:AUTO ON|OFF|1|0
:INPut[1]:ATTenuation:AUTO?
Example
INP:ATT:AUTO ON
Remote Command Notes
FSP, FSU
Preset
ON
State Saved
Saved in instrument state.
Remote Command
:INPut[1]:EATT <rel_ampl>
:INPut[1]:EATT?
Example
INP:EATT 15dB
Remote Command Notes
FSP, FSU
Preset
0
State Saved
Saved in instrument state.
Min
0