134
Compatible SCPI Commands
Supported SCPI Commands
SCPILC mode supports the range of –81.90 to 81.90 dB, not –200 to 200 dB as FSP.
[SENSe<1|2>:]CORRection:TRANsducer:SELect <name>
This command controls the transducer factor designated by <name>. If <name> does not exist yet, a new
transducer factor is created.
Currently, this command support 6 correction settings.
[SENSe<1|2>:]CORRection:TRANsducer:UNIT <string>
Specifies the unit of the transducer factor selected.
This command won’t invoke any action and the default value in the system is always “DB”.
Remote Command
[:SENSe]:CORRection:EGAin:INPut[:MAGNitude] <rel_ampl>
[:SENSe]:CORRection:EGAin:INPut[:MAGNitude]?
Example
CORR:EGA:INP 10DB
Remote Command Notes
FSP
Preset
0 dB
State Saved
Saved in instrument state.
Min
–81.90 dB
Max
81.90 dB
Backwards Compatibility SCPI
:SENSe1:CORRection:EGAin:INPut[:MAGNitude]
Remote Command
[:SENSe]:CORRection:TRANsducer:SELect <string>
[:SENSe]:CORRection:TRANsducer:SELect?
Example
CORR:TRAN:SEL “A”
CORR:TRAN:SEL?
Notes
Altogether there are only 6 different transducers can be selected. If there is a
new transducer, the most old one will be replaced by the new one.
Name of the transducer factor in string data form with a maximum of 8
characters.
FSU, FSP, FSE
State Saved
Saved in instrument state.
Remote Command
[:SENSe]:CORRection:TRANsducer:UNIT <string>
[:SENSe]:CORRection:TRANsducer:UNIT?
Example
CORR:TRAN:UNIT “DB”
CORR:TRAN:UNIT?