98
Compatible SCPI Commands
Supported SCPI Commands
When this function is activated, the APD measurement is turned off.
The query returns the current state of this setting.
CALCulate:STATistics:CCDF:X<1…3>? P0_01 | P0_1 | P1 | P10
Queries the level values for the probabilities 0.01%, 0.1%, 1% and 10%.
CALCulate:STATistics:NSAMples <100…1E9>
The command specifies the number of measurement points to be acquired for the statistical measurement
functions.
The minimum value supported by this mode is 1000.
The query returns the current value of this setting.
Remote Command
:CALCulate:STATistics:CCDF:STATe ON|OFF|1|0
:CALCulate:STATistics:CCDF:STATe?
Example
CALC:STAT:CCDF:STAT ON
Remote Command Notes
FSP, FSU
Preset
OFF
State Saved
Saved in instrument state.
Remote Command
CALCulate:STATistics:CCDF:X[1]|2|3? P0_01|P0_1|P1|P10
Example
CALC:STAT:CCDF:X? P10
Remote Command Notes
FSU
Remote Command
:CALCulate[1]:STATistics:NSAMples <int>
:CALCulate[1]:STATistics:NSAMples?
Example
CALC:STAT:NSAM 100000
CALC:STAT:NSAM?
Remote Command Notes
FSP, FSU
Preset
100000
State Saved
Saved in instrument state.
Min
1000
Max
1E9
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html