70
Compatible SCPI Commands
Supported SCPI Commands
The query returns the current value of this string.
CALCulate<1|2>:LIMit<1...6>:COPY 1 to 6 | <name>
Copies the limit line specified by the numeric suffix of LIMit onto the limit line specified by the
command’s parameter.
This mode does not support the parameter option <name>.
CALCulate<1|2>:LIMit<1...6>:NAME <name of limit line>
The command assigns a name to the limit line specified by the numeric suffix of LIMit. A limit line with
this name is created, if it does not already exist.
The query returns the current value of this string.
Remote Command
:CALCulate[1]:LIMit[1]|2|3|4|5|6:COMMent <string>
:CALCulate[1]:LIMit[1]|2|3|4|5|6:COMMent?
Example
CALC:LIM:COMM ‘GSM2’
CALC:LIM:COMM?
Remote Command Notes
FSE, FSP, FSU
Preset
“”
State Saved
Saved in instrument state.
Remote Command
:CALCulate[1]:LIMit[1]|2|3|4|5|6:COPY <int>
Example
CALC:LIM:COPY 2
Remote Command Notes
FSE, FSP, FSU
Min
1
Max
6
Remote Command
:CALCulate[1]:LIMit[1]|2|3|4|5|6:NAME <string>
:CALCulate[1]:LIMit[1]|2|3|4|5|6:NAME?
Example
CALC:LIM:NAME ‘test’
CALC:LIM:NAME?
Remote Command Notes
FSE, FSP, FSU
Preset
'REM1' to 'REM6' for lines 1 to 6
State Saved
Saved in instrument state.