112
Compatible SCPI Commands
Supported SCPI Commands
DISPlay[:WINDow<1|2>]:TRACe<1 to 3>:Y[:SCALe]:PDIVision
The command specifies the scaling of the Y-axis, using the currently-selected unit.
This mode supports only dB units.
The query returns the current value of this setting.
DISPlay[:WINDow<1|2>]:TRACe<1 to 3>:Y:SPACing LINear | LOGarithmic| LDB
The command switches between linear and logarithmic display.
This mode does not support the parameter option LDB.
The query returns the current state of this setting.
Remote Command Notes
FSE, FSP, FSU
Preset
0 dB
State Saved
Saved in instrument state.
Min
–200.0 dB
Max
200 dB
Backwards Compatibility SCPI
:DISPlay:TRACe[1]|2|3:Y[:SCALe]:RLEVel:OFFSet
Remote Command
DISPlay:WINDow[1]:TRACe<1 to 3>:Y[:SCALe]:PDIVision
<rel_ampl>
DISPlay:WINDow[1]:TRACe<1 to 3>:Y[:SCALe]:PDIVision?
Example
DISP:WIND:TRAC:Y:PDIV 5
DISP:WIND:TRAC:Y:PDIV?
Remote Command Notes
FSE, FSP, FSU
State Saved
Saved in instrument state.
Backwards Compatibility SCPI
:DISPlay[1]:TRACe[1]|2|3
:Y[:SCALe]:PDIVision
Remote Command
:DISPlay:WINDow[1]:TRACe[1]|2|3:Y:SPACing
LINear|LOGarithmic
:DISPlay:WINDow[1]:TRACe[1]|2|3:Y:SPACing?
Example
DISP:WIND:TRAC:Y:SPAC LOG
DISP:WIND:TRAC:Y:SPAC?
Remote Command Notes
FSE, FSP, FSU
Preset
LOG
State Saved
Saved in instrument state.
Backwards Compatibility SCPI
:DISPlay[1]:TRACe[1]|2|3
:Y:SPACing
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html