30
Front Panel Operations
Mode Setup
Mode Setup
NOTE
License N9062A-2FP or W9062A-2FP is required for emulating FSE, FSP or
FSU.
Allows you to adjust the mode setup configuration for current mode.
Selecting Compatibility Model
To select the compatibility model you wish to emulate, press the
Mode Setup
hardkey on the front panel
and then press
Compatibility Model
.
Figure 2-3
shows the menu map that allows you to select the
Compatibility Mode from FSE, FSP and FSU.
Figure 2-3
Mode Setup > Compatibility Mode Selection Menu Map
NOTE
When the
Mode IDN Response
selects Compatibility Model, it does not affect the
response to the command “ID?” which always returns the model number of your
X-Series analyzer, like "N9020A".
The selected compatibility model determines the response to the “IDN?”
command. Setting the Compatibility Model to ‘R&S FSE’, ‘R&S FSP’ or ‘R&S
FSU’ affects the response to the SCPI command ‘*IDN?’ This command will
return the model number and firmware version number of the corresponding
instrument.
Key Path:
Front-panel key
Initial S/W Revision:
A.06.00 or later
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html