163
Compatible SCPI Commands
Supported SCPI Commands
This mode supports the level range:–200 to 100 dBm.
TRIGger<1|2>[:SEQuence]:LEVel:VIDeo 0 to 100PCT
Sets the video trigger level.
TRIGger<1|2>[:SEQuence]:HOLDoff
Specifies the trigger delay period.
Remote Command
:TRIGger[1][:SEQuence]:LEVel:RFPower <real>
:TRIGger[1][:SEQuence]:LEVel:RFPower?
Example
TRIG:LEV:RFP –20DBM
Remote Command Notes
FSP
Preset
–20dBm
State Saved
Saved in instrument state.
Min
–200dBm
Max
100dBm
Remote Command
TRIGger[1][:SEQuence]:LEVel:VIDeo <real>
TRIGger[1][:SEQuence]:LEVel:VIDeo?
Example
TRIG:LEV:VID 40
TRIG:LEV:VID?
Remote Command Notes
FSE, FSP, FSU
Preset
50
State Saved
Saved in instrument state.
Min
0
Max
100
Remote Command
:TRIGger[1][:SEQuence]:HOLDoff <time>
:TRIGger[1][:SEQuence]:HOLDoff?
Example
TRIG:HOLD 100ms
Remote Command Notes
FSE, FSP, FSU
Preset
0s
State Saved
Saved in instrument state.
Min
0s
Max
0.5s