91
Compatible SCPI Commands
Supported SCPI Commands
This command is an event, and therefore has no *RST value and no query.
CALCulate<1|2>:MARKer<1...4>:MAXimum:AUTO ON | OFF
For marker 1, the command switches on or off operation of automatic maximum peak search at the end
of each sweep.
The marker search limit settings are taken into account.
For this command, the numeric suffix <1 to 4> is ignored.
The query returns the current state of this setting.
CALCulate<1|2>:MARKer<1...4>:MINimum:AUTO ON | OFF
For marker 1, the command switches on or off operation of automatic minimum peak search at the end of
each sweep.
The marker search limit settings are taken into account.
For this command, the numeric suffix <1 to 4> is ignored.
The query returns the current state of this setting.
Remote Command
CALCulate[1]:MARKer[1]|2|3|4:MINimum:RIGHt
Example
CALC:MARK:
MIN
:RIGH
Remote Command Notes
FSE, FSP, FSU
Remote Command
:CALCulate[1]:MARKer[1]|2|3|4:MAXimum:AUTO ON|OFF|1|0
:CALCulate[1]: MARKer[1]|2|3|4:MAXimum:AUTO?
Example
CALC:MARK:MAX:AUTO ON
CALC:MARK:MAX:AUTO?
Remote Command Notes
FSU
Preset
OFF
State Saved
Saved in instrument state.
Remote Command
:CALCulate[1]:MARKer[1]|2|3|4:MINimum:AUTO ON|OFF|1|0
:CALCulate[1]: MARKer[1]|2|3|4:MINimum:AUTO?
Example
CALC:MARK:MIN:AUTO ON
CALC:MARK:MIN:AUTO?
Remote Command Notes
FSU
Preset
OFF
State Saved
Saved in instrument state.