97
Compatible SCPI Commands
Supported SCPI Commands
This mode supports ACPower, CPOWer, MCACpower, OBANdwidth and OBWidth.
CALCulate<1|2>:MARKer<1...4>:FUNCtion:POWer:RESult:PHZ ON | OFF
The command switches the query response for power measurement results between absolute values
(OFF) and output referred to measurement bandwidth (ON).
The query returns the current state of this setting.
CALCulate:STATistics Subsystem
The CALCulate:STATistics subsystem controls the statistical measurement functionality.
CALCulate:STATistics:APD[:STATe] ON | OFF
This command is accepted, but takes no action and reports no error.
CALCulate:STATistics:CCDF[:STATe] ON | OFF
The command activates or deactivates measurement of the complementary cumulative distribution
function (CCDF).
Remote Command
CALCulate[1]:MARKer[1]|2|3|4:FUNCtion:POWer:RESult?
ACPower|CPOWer|MCACpower|OBANdwidth|OBWidth|CN|CN0
Example
CALC:MARK:FUNC:POW:RES? OBW
Remote Command Notes
FSE, FSP, FSU
Remote Command
:CALCulate[1]:MARKer[1]|2|3|4:FUNCtion:POWer:RESult:PHZ
ON|OFF|1|0
:CALCulate[1]:MARKer[1]|2|3|4:FUNCtion:POWer:RESult:PHZ
?
Example
CALC:MARK:FUNC:POW:RES:PHZ OFF
CALC:MARK:FUNC:POW:RES:PHZ?
Remote Command Notes
FSP, FSU
Preset
OFF
State Saved
Saved in instrument state.
Remote Command
:CALCulate:STATistics:APD:STATe ON|OFF|1|0
:CALCulate:STATistics:APD:STATe?
Example
CALC:STAT:APD:STAT ON
Remote Command Notes
FSP, FSU
Preset
OFF
State Saved
Saved in instrument state.
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html