Compatible SCPI Commands
Supported SCPI Commands
STATus:QUEStionable:PTRansition
This mode supports the maximum value of 32767, not 65535 as in FSx.
STATus:QUEStionable:NTRansition
Specifies the edge detectors of all bits in the STATus:OPERation register from 1 to 0 for the transitions
of the CONDition bit.
This mode supports the maximum value of 32767, not 65535 as FSx.
STATus:QUEStionable:FREQuency[:EVENt]
Queries the contents of the EVENt section in the STATus:QUEStionable: FREQuency register.
STATus:QUEStionable:FREQuency:CONDition
Queries the contents of the CONDition section in the
STATus:QUEStionable:FREQuency register. The contents of the CONDition section are not deleted
Remote Command
:STATus:QUEStionable:PTRansition <integer>.
:STATus:QUEStionable:PTRansition?
Example
STAT:QUES:PNTR 16
Remote Command Notes
FSE, FSP, FSU
Preset
0
Min
0
Max
32767
Remote Command
:STATus:QUEStionable:NTRansition <integer>.
:STATus:QUEStionable:NTRansition?
Example
STAT:QUES:NTR 16
Remote Command Notes
FSE, FSP, FSU
Preset
0
Min
0
Max
32767
Remote Command
:STATus:QUEStionable:FREQuency[:EVENt]?
Example
STAT:QUES:FREQ?
Remote Command Notes
FSE, FSP, FSU
Preset
0