153
Compatible SCPI Commands
Supported SCPI Commands
STATus:OPERation:PTRansition
Specifies the edge detectors of all bits in the STATus:OPERation register from 0 to 1 for the transitions
of the CONDition bit.
This mode supports the maximum value of 32767, not 65535 as FSx.
STATus:OPERation:NTRansition
Specifies the edge detectors of all bits in the STATus:OPERation register from 1 to 0 for the transitions
of the CONDition bit.
This mode supports the maximum value of 32767, not 65535 as in FSx.
STATus:PRESet
Resets the edge detectors and ENABle parts of all registers to a defined value.
Remote Command
:STATus:OPERation:PTRansition <integer>
:STATus:OPERation:PTRansition?
Example
STAT:OPER:PTR 1
Remote Command Notes
FSE, FSP, FSU
Preset
32767
Min
0
Max
32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode
All
Remote Command
:STATus:OPERation:NTRansition <integer>
:STATus:OPERation:NTRansition?
Example
STAT:OPER:NTR 1
Remote Command Notes
FSE, FSP, FSU
Preset
0
Min
0
Max
32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Remote Command
:STATus:PRESet
Example
STAT:PRES