Compatible SCPI Commands
Agilent N9062A/W9062A Supported IEEE SCPI Commands
*RST
This command presets the instrument to a factory defined condition that is appropriate for
remote programming operation. *RST is equivalent to performing the commands:
1.:SYSTem:PRESet, with preset type set to MODE.
2.*CLS which clears the STATus bits and error queue
3.:INITiate:CONTinous OFF which selects single sweep/measurement
*RST does not change the mode and only resets the parameters for the current mode.
*SAV <register>
This command saves the instrument state to the specified instrument memory register.
*SRE <integer>
*SRE?
This command sets the value of the service request enable register.
The query returns the value of the register.
*STB?
Returns the value of the status byte register without erasing its contents.
*TRG
This command triggers the instrument. Use the command :TRIGger[:SEQuence]:SOURce
to select the trigger source.
*TST?
This query performs a self test and returns a number indicating the success of the testing.
A zero is returned if the test is successful.
*WAI
This command causes the instrument to wait until all pending commands/processes are
completed before executing any additional commands.
There is no query form for the command.
Table 3-2
IEEE 488.2 SCPI Commands
SCPI Commands
Description