72
Compatible SCPI Commands
Supported SCPI Commands
CALCulate<1|2>:LIMit<1...6>:ESPectrum:RESTore
Restores the standard limit lines for the spectrum emission mask measurement. Any modifications that
were made to the standard limit lines are lost, and the default setting is restored.
This command is accepted, but takes no action and reports no error.
CALCulate:LIMit:ACPower Subsystem
The CALCulate:LIMit:ACPower subsystem checks the limit for adjacent channel power (ACP)
measurement.
CALCulate<1|2>:LIMit<1…6>:ACPower[:STATe]
Switches on and off the limit check for ACP measurements in the selected measurement window. In
addition, you should use the commands CALCulate:LIMit:ACPower:ACHannel:STATe or
CALCulate:LIMit:ACPower:ALTernate:STATe to specify whether the limit check is to be performed for
the upper/lower adjacent channel or for the alternate adjacent channels.
The numeric suffix is irrelevant for this command.
CALCulate<1|2>:LIMit<1…6>:ACPower:ACHannel[:RELative]
Specifies the relative limit of the upper/lower adjacent channel for ACP measurements in the selected
measurement window. The reference value for the relative limit value is the measured channel power.
The numeric suffix is irrelevant for this command.
Remote Command
:CALCulate[1]:LIMit[1]|2|3|4|5|6:ESPectrum:RESTore
Example
CALC:LIM:ESP:REST
Remote Command Notes
FSP
Remote Command
:CALCulate[1]:LIMit[1]|2|3|4|5|6:ACPower[:STATe]
ON|OFF|1|0
:CALCulate[1]:LIMit[1]|2|3|4|5|6:ACPower[:STATe]?
Example
CALC:LIM:ACP ON
CALC:LIM:ACP?
Notes
FSU, FSP, FSE
Preset
OFF
State Saved
Saved in instrument state.
Remote Command
:CALCulate[1]:LIMit[1]|2|3|4|5|6:ACPower:ACHannel[:RELa
tive] <real>,<real>
:CALCulate[1]:LIMit[1]|2|3|4|5|6:ACPower:ACHannel[:RELa
tive]?