77
Compatible SCPI Commands
Supported SCPI Commands
CALCulate<1|2>:LIMit<1…6>:ACPower:ALTernate<1|2>:RESult?
Queries the result of the limit check for the first/second alternate adjacent channel in the selected
measurement window for ACP measurements.
The numeric suffix after ALTernate<1|2> denotes the first or the second alternate channel. The numeric
suffixes <1 to 6> are irrelevant for this command.
If the power measurement of the adjacent channel is switched off, the command produces a query error.
The result is returned in the form <result>, <result> where <result> = PASSED | FAILED and where the
first (second) returned value denotes the lower (upper) alternate adjacent channel.
CALCulate:LIMit:CONTrol Subsystem
The CALCulate:LIMit:CONTrol subsystem controls the X-axis.
NOTE
All limit functions are not supported in the CCDF and APD measurements.
CALCulate<1|2>:LIMit<1...6>:CONTrol[:DATA] <numeric_value>,<numeric_value>..
The command specifies the X-axis values (frequencies or times) of the upper or lower limit lines.
This mode does not support any unit for this command.
The query returns the current values of these settings.
CALCulate<1|2>:LIMit<1...6>:CONTrol:DOMain FREQuency | TIME
The command switches between frequency or time domain for the X-axis.
Remote Command
:CALCulate[1]:LIMit[1]|2|3|4|5|6:ACPower:ALTernate[1]|2
:RESult?
Example
CALC:LIM:ACP:ALT:RES?
CALC:LIM:ACP:ALT2:RES?
Notes
FSU, FSP, FSE
Remote Command
:CALCulate[1]:LIMit[1]|2|3|4|5|6:CONTrol[:DATA]
<real>,<real>…
:CALCulate[1]:LIMit[1]|2|3|4|5|6:CONTrol[:DATA]?
Example
CALC:LIM:CONT:DATA 1.0e7,2.0e7,1.0e8,3.0e8,1.0e9
CALC:LIM:CONT:DATA?
Remote Command Notes
FSE, FSP, FSU
State Saved
Saved in instrument state.