78
Compatible SCPI Commands
Supported SCPI Commands
The query returns the current state of this setting.
CALCulate<1|2>:LIMit<1...6>:CONTrol:OFFSet <numeric_value>
The command defines an offset for the X-axis value of the specified relative limit line, in the frequency
or time domain.
The query returns the current value of this setting.
CALCulate<1|2>:LIMit<1...6>:CONTrol:MODE RELative | ABSolute
The command switches between relative or absolute scaling for the X-axis of the specified limit line.
The query returns the current state of this setting.
Remote Command
:CALCulate[1]:LIMit[1]|2|3|4|5|6:CONTrol:DOMain
FREQuency|TIME
:CALCulate[1]:LIMit[1]|2|3|4|5|6:CONTrol:DOMain?
Example
CALC:LIM:CONT:DOM FREQ
CALC:LIM:CONT:DOM?
Remote Command Notes
FSE, FSP, FSU
Preset
FREQuency
State Saved
Saved in instrument state.
Range
FREQuency | TIME
Remote Command
:CALCulate[1]:LIMit[1]|2|3|4|5|6:CONTrol:OFFSet <real>
:CALCulate[1]:LIMit[1]|2|3|4|5|6:CONTrol:OFFSet?
Example
CALC:LIM:CONT:OFFS 1
CALC:LIM:CONT:OFFS?
Remote Command Notes
FSE, FSP, FSU
Preset
0
Remote Command
:CALCulate[1]:LIMit[1]|2|3|4|5|6:CONTrol:MODE
RELative|ABSolute
:CALCulate[1]:LIMit[1]|2|3|4|5|6:CONTrol:MODE?
Example
CALC:LIM:CONT:MODE ABS
CALC:LIM:CONT:MODE
Remote Command Notes
FSE, FSP, FSU
Preset
ABS
State Saved
Saved in instrument state.