![Agilent Technologies CXA N9000A Скачать руководство пользователя страница 143](http://html.mh-extra.com/html/agilent-technologies/cxa-n9000a/cxa-n9000a_user-manual_2868013143.webp)
142
Compatible SCPI Commands
Supported SCPI Commands
[SENSe<1|2>:]POWer:ACHannel:SPACing:ACHannel 100 Hz to 2000 MHz
Specifies the channel spacing of the adjacent channel to the TX channel. At the same time, the spacing of
alternate adjacent channels 1 and 2 is set to double or triple the entered value.
This command is available only in the frequency domain (span > 0).
[SENSe<1|2>:]POWer:ACHannel:SPACing:ALTernate<1|2> 100 Hz to 2000 MHz
Specifies the spacing between the first (ALTernate1) or second alternate adjacent channel (ALTernate2)
and the TX channel.
[SENSe<1|2>:]POWer:ACHannel:TXCHannel:COUNt 1 | 2 | 3 | 4
Specifies the number of carrier signals. This command is only available for multicarrier channel and
Remote Command Notes
FSE, FSP, FSU
Preset
20 kHz
State Saved
Saved in instrument state.
Min
100 Hz
Max
2 GHz
Remote Command
[:SENSe]:POWer:ACHannel:SPACing:ACHannel <freq>
[:SENSe]:POWer:ACHannel:SPACing:ACHannel?
Example
POW:ACH:SPAC:ACH 33kHz
Remote Command Notes
FSE, FSP, FSU
Preset
20KHz
State Saved
Saved in instrument state.
Min
100Hz
Max
2000MHz
Remote Command
[:SENSe]:POWer:ACHannel:SPACing:ALTernate[1]|2 <freq>
[:SENSe]:POWer:ACHannel:SPACing:ALTernate[1]|2?
Example
POW:ACH:SPAC:ALT1 100kHz
Remote Command Notes
FSE, FSP, FSU
Preset
40000|60000
State Saved
Saved in instrument state.
Min
100Hz
Max
2000MHz
This Manual:http://www.manuallib.com/agilent/n9062a-signal-analyzer-user-guide.html