51
3 Installing BenchCel Microplate Handler
Connecting the pendant directly to the BenchCel device
BenchCel Microplate Handler User Guide
Connecting the pendant directly to the BenchCel device
About this topic
If you are integrating the BenchCel Microplate Handler with devices that are not
equipped with safety interlock circuitry, for example, the PlateLoc Sealer or the
Microplate Labeler, use the procedure in this topic to connect the emergency-stop
pendant.
Figure
Emergency-stop pendant
The pendant must be connected to close the safety interlock circuit. The circuit must
be closed for the device to operate.
IMPORTANT
If the BenchCel Microplate Handler is integrated with devices that are equipped with
safety interlock circuitry, such as the Bravo Platform, you connect the pendant through
a main emergency-stop control, such as the Robot Disable Hub. See
pendant through the Robot Disable Hub” on page 53
.
Setting up the BenchCel Microplate
Handler
Operating the BenchCel Microplate
Handler
“Workflow for operating the BenchCel
Microplate Handler” on page 116
For more information about…
See…