![Agilent Technologies BenchCel Скачать руководство пользователя страница 179](http://html1.mh-extra.com/html/agilent-technologies/benchcel/benchcel_user-manual_2867698179.webp)
169
A Quick reference
BenchCel Diagnostics - Controls tab
BenchCel Microplate Handler User Guide
Pick from <teachpoint>
Picks up the plate from the selected teachpoint, and
then the robot backs away from the teachpoint into
the safe zone.
Place at <teachpoint>
Moves the robot from its current position and places
the plate at the selected teachpoint. After placing the
plate, the robot backs away from the teachpoint into
the safe zone.
Transfer to <teachpoint>
Moves a plate from the currently selected teachpoint
to the other selected teachpoint.
To transfer a plate from a stack, you must first click
Load Stacker. During the transfer, the robot
downstacks the plate, places it at the other selected
teachpoint, and then backs into the safe zone.
If the robot is transferring a plate from one
teachpoint to another, the robot picks up the plate
from the currently selected teachpoint, places the
plate at the other selected teachpoint, and then backs
into the safe zone.
Delid from <teachpoint>
Lidded labware only.
Removes the lid from the
labware that is at the selected teachpoint.
Relid to <teachpoint>
Lidded labware only.
Replaces the lid on the labware
that is at the selected teachpoint.
Control or command