173
A Quick reference
BenchCel Diagnostics - Labware tab
BenchCel Microplate Handler User Guide
BenchCel Diagnostics - Labware tab
You can use the Labware tab to perform the following procedures:
•
“Setting sensor thresholds” on page 99
•
“Changing labware parameters” on page 154
The settings in the Labware tab are also available in the Labware Editor.
Commands
Plate Dimensions area
Command
Description
Apply/Save Labware
Saves the updated settings in the Labware tab.
Parameters
Description
Thickness
The distance (mm) from the bottom surface of the
plate to the top surface of the plate.
For a tip box, this is the distance from the bottom
surface of the tip box to the top of the tips.
To increase the number of contact points, measure
the distance at the corner of the plate or tip box (using
calipers). This method is especially useful if the plate
has a lip at the top which could cause the caliper to
angle inward, producing inaccurate measurements.