7 Maintenance and troubleshooting
Replacing the fuse
134
BenchCel Microplate Handler User Guide
Replacing the fuse
About this topic
If the main fuse is bad, the BenchCel Microplate Handler will not turn on. This topic
describes how to replace the main fuse in the device.
Before you start
CAUTION
A blown fuse can indicate more serious problems. If the new fuse blows after
replacement, contact Agilent Technical Support.
CAUTION
Using an incorrect fuse can damage the BenchCel Microplate Handler.
Use only the specified fuse type: 5 A, 250 V, 5 mm x 20 mm, fast acting. You can order
fuses from Agilent Technologies.
On the BenchCel control panel, verify the power rating (120 V or 230 V) for your region
matches the display on the fuse enclosure.
Procedure
To replace the fuse in the power switch:
1
Shut down the BenchCel Microplate Handler, and unplug the power cable from the
rear panel connector.
2
At the rear panel power switch enclosure, use a small flat-head screwdriver
(2.5 mm) to pry open the tab on the fuse enclosure and open the enclosure cover.