7 Maintenance and troubleshooting
Software error messages
140
BenchCel Microplate Handler User Guide
Deadlock detected
The microplate is assigned to the
wrong labware class.
Check that the labware belongs to the correct
labware class in the Labware Editor. Most
microplates should belong to the Uses
Standard Plate Pad class. If you created a
special class for a particular microplate, make
sure it belongs to the special class.
For more information about setting up
labware classes, see the
VWorks Automation
Control Setup Guide
.
The protocol run is unable to
continue. Some or all of the target
locations are unavailable, so the
microplates cannot be moved to
those locations.
Review the protocol and the run log. Run the
simulator to determine the reason for the
deadlock. In addition, check that the number
of simultaneous microplates is not too high.
See the
VWorks Automation Control User
Guide
.
The device was not
found
The BenchCel device is not
connected to the controlling
computer, or the incorrect profile is
used.
To connect the device:
1
Turn off the BenchCel Microplate Handler
and exit the VWorks software.
2
Ensure the BenchCel Microplate Handler
is connected to the controlling computer.
See
.
3
Wait 10 seconds, and then turn on the
BenchCel Microplate Handler.
4
Wait for the BenchCel Microplate Handler
to finish its homing routine, and then start
the VWorks software.
To check the profile:
1
In BenchCel Diagnostics, click the
Profiles
tab.
2
Check or select the correct profile. Verify
that the profile uses the correct the COM
port number (serial) or Ethernet ID.
Flash operation not
successful
The robot could not write to its
flash memory.
In the error message dialog box, click
Retry
. If
clicking Retry does not resolve the problem,
restart the BenchCel Microplate Handler.
If the problem persists, contact Agilent
Technical Support.
The gripper positions
are too close
The Gripper open position
parameter value is less than the
Gripper holding plate position
parameter value.
In the labware definition, make sure the
Gripper open position value is 3 mm or greater
than the Gripper holding plate position value.
You can change the parameters in the
Labware Editor or in BenchCel Diagnostics.
Error message
Cause
Solution