BenchCel Diagnostics - Jog/Teach tab
BenchCel Microplate Handler User Guide
Robot movement area
Plate presence threshold
The value that determines if a microplate is detected.
Any value equal to or greater than the threshold
means a microplate is present. A lesser value means
there is no microplate.
Note:
The setting is also available in the Labware tab.
Apply & save
The command that saves the updated settings in the
Jog/Teach tab.
Command
Description
CCW
Jogs the robot arm counterclockwise from the
current position by the specified
theta
-axis increment.
CW
Jogs the robot arm clockwise from the current
position by the specified
theta
-axis increment.
Left
Jogs the robot head left from the current position by
the specified
x
-axis increment.
Up
Jogs the robot head up from the current position by
the specified
z
-axis increment.
Right
Jogs the robot head right from the current position by
the specified
x
-axis increment.
Down
Jogs the robot head down from the current position
by the specified
z
-axis increment.
Gripper
Open
Opens the robot grippers by the specified grip
increment.
Close
Closes the robot grippers by the specified grip
increment.
Full open
Opens the robot grippers to the robot Gripper open
position set in the Labware tab.
Full close
Closes the robot grippers to the robot Gripper holding
plate position set in the Labware tab.
Go Home
Moves the robot head to the home position, which is
at the center of the BenchCel Microplate Handler, and
the robot arms are perpendicular to the
x
-axis. You
send the robot to the home position if you want the
robot out of the way in a safe position.