![Agilent Technologies BenchCel Скачать руководство пользователя страница 145](http://html1.mh-extra.com/html/agilent-technologies/benchcel/benchcel_user-manual_2867698145.webp)
135
7 Maintenance and troubleshooting
Replacing the fuse
BenchCel Microplate Handler User Guide
Figure
Fuse enclosure location
3
At the top of the enclosure, insert the screwdriver head in the notch to dislodge the
red fuse cartridge. Slide the fuse cartridge all the way out of the enclosure.
Figure
Opened fuse enclosure
4
Inspect the fuses on each side of the cartridge, and then replace the suspect fuse
or fuses for the type of system you are using (120 V or 230 V power).
Figure
Fuse cartridge (A) Right side with correctly installed fuse, B) Left side with spare
fuse
IMPORTANT
Ensure the fuse is installed correctly in the cartridge. Otherwise, the device might not
turn on.
5
Slide the fuse cartridge back into the fuse enclosure.