6 Preparing for a run
Filling and emptying the labware racks
BenchCel Microplate Handler User Guide
IMPORTANT
If you are removing labware from a mounted rack, ensure that the stack is unloaded.
To unload the stack, open BenchCel Diagnostics. On the Control page, click Stacker in
the graphical display area, and then choose Unload Plates. Alternatively, you can click
Unload All to unload all the stacks.
2
Carefully slide the labware in small stacks, up and out of the top of the rack.
Filling a front-load rack
The doors on the front-load rack provide easy access for placing labware into the front
of a rack that is mounted on a device.
Figure
Door mechanism on the front-load rack
To place labware into a front-load rack:
1
If the rack is mounted on the BenchCel device, verify that the stack is in the
unloaded state.
To unload the stack, open BenchCel Diagnostics. On the
Control
page, click
Stacker
,
and then choose
Unload Plates
. Alternatively, you can click
Unload All
to unload all
the stacks.
2
On each side of the rack, slide the Door-release (black) buttons forward, while
pushing outward on the thumb tabs. The rack doors open.
Figure
Opening the front-load rack
Door-release
button
Door
Door
Thumb tab
Stacker gripper