5 Setting sensor thresholds
Calculating the Plate presence threshold
BenchCel Microplate Handler User Guide
2
In
BenchCel Diagnostics
on the
Controls
page, click the
Labware
tab. In the
Sensors
area, set the initial values as follows:
•
Plate presence
threshold
at 225 units.
•
Intensity
at 100%.
Click
Apply and save labware parameters
.
VWorks Plus only
. If an audit trail is being logged, the Audit Comment dialog box
opens. Select or type the audit comment, and then click
OK
.
3
Click the
Jog/Teach
tab. In the
Stacker Sensors
area:
a
Select stack 1 from the list.
b
Record the
Plate presence
reading.
c
Repeat steps a and b for each of the remaining stacks.
d
In the list, select the number of the stack that has the highest Plate presence
reading.
4
On the
Labware
tab, adjust the
Intensity
value so that the highest
Plate presence
reading (Jog/Teach tab) is less than or equal to 175. Click
Apply and save labware
parameters
.
VWorks Plus only
. If an audit trail is being logged, the Audit Comment dialog box
opens. Select or type the audit comment, and then click
OK
.
Use the following guidelines when adjusting the Intensity value:
•
For the most reliable sensor function, use the highest Intensity value possible
while keeping the Plate presence reading no greater than 175.
•
If the Plate presence reading is greater than 175, decrease the Intensity value
incrementally so that the Plate presence reading is reduced to 175.
For example, if a light-colored microplate has a Plate presence reading of 200,
decreasing the Intensity value to 20 reduces the reading to 175.