7 Maintenance and troubleshooting
Cleaning up after a protocol run
132
BenchCel Microplate Handler User Guide
Related information
Cleaning up after a protocol run
About this topic
This topic describes the tasks you should perform when you have finished a protocol
run.
Procedure
CAUTION
Make sure you clean up spilled liquids immediately. Do not use harsh abrasives,
corrosive cleaning agents, or metal brushes to clean any BenchCel component or
accessory.
To clean up after a run:
1
Follow the VWorks prompts for post-run software procedures, such as unloading
the microplates. See the
VWorks Automation Control User Guide
for detailed
instructions.
2
Unload used sample microplates from the labware racks. See
the labware racks” on page 119
3
Remove the labware racks from the BenchCel Microplate Handler. See
and uninstalling the labware racks” on page 123
.
IMPORTANT
Make sure the power and compressed air are turned on when removing the racks.
4
Remove manually placed microplates from platepads and external devices.
5
Use a clean soft cloth or laboratory wipes to clean up spills on any part of the
BenchCel device immediately after a protocol run.
If necessary, use standard laboratory wipes and a mild detergent or isopropyl
alcohol to clean the exterior painted white surfaces and the metal surfaces of dust,
grime, chemical deposits, and other debris.
6
Make sure the
x
-axis tracks are free of debris.
For information about...
See...
Cleanup procedure after every
protocol run
“Cleaning up after a protocol run” on
page 132
Air source requirements
“Compressed air requirements” on
page 39
Checking and adjusting the
stacker gripper pressure (internal
air pressure)