7 Maintenance and troubleshooting
Routine maintenance
130
BenchCel Microplate Handler User Guide
Routine maintenance
About this topic
This topic provides recommendations for maintaining the BenchCel Microplate
Handler.
Cleaning up spills and periodic cleaning
In general, practice good housekeeping by cleaning up spills and routinely cleaning after
using the BenchCel Microplate Handler.
Use a clean soft cloth or laboratory wipes to clean up spills on any part of the device
immediately after a protocol run.
Clean the BenchCel device weekly or as needed. Use standard laboratory wipes and a
mild detergent or isopropyl alcohol to clean the exterior painted white surfaces and the
metal surfaces of dust, grime, chemical deposits, and other debris.
Also see
“Cleaning up after a protocol run” on page 132
.
Monthly inspection and maintenance
Every month, check the following:
•
Robot gripper pins
. Make sure they are not too dull to grip the microplates. If the
pins have become dull, contact Agilent Technical Support to replace them.
Figure
Robot head (side view)
Check the pins on each
robot gripper.