Features and Functions
2
34980A User’s Guide
17
Measurement Functions
The following table shows which DMM measurement functions are
supported by each of the multiplexer modules.
Note that similar considerations must be taken into account on the
34931A, 34932A, and 34933A matrix modules. Since the matrix modules
cannot be incorporated into a scan list, you must use the Stand- Alone
DMM Mode for these modules.
Front Panel Operation:
DMM or Channel (Configure) > DMM MEASUREMENT
Use the knob (or numeric keypad) to select the desired channel. Then
select the desired measurement function for this channel. You are
automatically guided to the next level of the menu where you can
configure other measurement parameters (range, integration time, etc.).
Remote Interface Operation:
You can select the measurement function using
the
CONFigure
and
MEASure?
commands. For example, the following
command configures the specified channel for dc voltage measurements.
CONF:VOLT:DC 10,DEF,(@3001)
Function
34921A
40-Ch Arm
MUX
34922A
70-Ch Arm
MUX
34923A
40-Ch Reed
MUX
(2-Wire)
34923A
80-Ch Reed
MUX
(1-Wire)
34924A
70-Ch Reed
MUX
34925A
40-Ch FET
MUX
(2-Wire)
34925A
80-Ch FET
MUX
(1-Wire)
Voltage, AC/DC
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Current, AC/DC
Yes
1
No
No
No
No
No
No
Frequency/Period
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Ohms 2-Wire
Yes
Yes
Yes
5
Yes
5
Yes
5
Yes
6
Yes
6
Ohms 4-Wire
Yes
Yes
Yes
5
No
Yes
5
Yes
6
No
Thermocouple Yes
2
Yes
3
Yes
3,4
Yes
3,4
Yes
3,4
Yes
3
Yes
3
RTD 2-Wire
Yes
Yes
Yes
5
Yes
5
Yes
5
No
No
RTD 4-Wire
Yes
Yes
Yes
5
No
Yes
5
Yes
6
No
Thermistor Yes
Yes
Yes
5
Yes
5
Yes
5
No
No
1
Direct current measurements are allowed on channels 41 through 44 only (for all other channels, external shunts are required).
2
Optional 34921T Terminal Block is required for thermocouple measurements with built-in internal reference junction.
3
A fixed or external reference junction temperature is required for thermocouple measurement with this module.
4
Impact of higher offset voltage specification (< 50
µ
V) must be taken into consideration.
5
1 k
Ω
or higher range used unless 100
Ω
series resistors are bypassed on module.
6
10 k
Ω
or higher range used for loads over approximately 300
Ω
due to series resistance of FET channels.
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...