Features and Functions
2
34980A User’s Guide
19
Measurement Resolution
Resolution is expressed in
number of digits
the internal DMM can
measure or display on the front panel. You can set the resolution to 4, 5,
or 6 full digits, plus a “½” digit which can be “0” or “1”. To increase the
measurement accuracy and improve noise rejection, select 6½ digits.
To increase the measurement speed, select 4½ digits.
•
For ac voltage measurements, the resolution is fixed at 6½ digits.
The only way to control the reading rate for ac measurements is by
changing the channel delay (see
page 56
) or by setting the ac filter to
the highest frequency limit (see
page 37
).
•
The specified resolution is used for all measurements on the selected
channel. If you have applied Mx+B scaling or have assigned alarms to
the selected channel, those measurements are also made using the
specified resolution. Measurements taken during the Monitor function
also use the specified resolution.
•
Changing the number of digits does more than just change the
resolution of the instrument. It also changes the
integration time
,
which is the period the instrument’s analog- to- digital (A/D) converter
samples the input signal for a measurement. See
“Custom A/D
Integration Time”
on page 20 for more information.
•
The
CONFigure
and
MEASure?
commands contain an optional parameter
which allows you to specify the resolution.
•
The instrument returns to 5½ digits when the measurement function is
changed and after a Factory Reset (
*RST
command). An Instrument
Preset (
SYSTem:PRESet
command) or Card Reset (
SYSTem:CPON
command) does
not
change the resolution setting.
Front Panel Operation:
DMM or Channel (Configure) > INTEGRATION > NPLC
First, select the measurement function on the active channel. You are
automatically guided to the next level of the menu where you can select a
specific resolution.
Remote Interface Operation:
Specify the resolution in the same units as
the measurement function, not in number of digits. For example, if the
function is dc voltage, specify the resolution in volts. For frequency,
specify the resolution in hertz.
You can select the resolution using parameters in the
CONFigure
and
MEASure?
commands. For example, the following command selects the
10 Vdc range with 4½ digits of resolution on the specified channel.
CONF:VOLT:DC 10,0.001,(@3001)
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...