278
34980A User’s Guide
10
64-Bit Digital I/O Module with Memory and Counter
Set the number of times to output the trace.
Each trace can be output once,
multiple times, or infinitely. The
SOURce:DIGital:MEMory:NCYCles
command sets the number of times to output the trace. If not set to
infinite, you can output the trace from 1 to 255 times (the output is
controlled by the handshake).
Load the trace(s) into memory.
Named traces are downloaded using the
TRACe:DATA:DIGital
command. The channel width used should match the
width of the channel set in step 1. If you change the width of a bank,
all traces in memory are cleared. The trace names must start with a
character and may be up to 12 characters in length. The trace name used
must be unique to the bank. Up to 32 traces may be downloaded (up to
the maximum memory).
Traces can be added or deleted only when memory is disabled. Memory
output cannot be enabled unless the bank has a trace assigned to it.
For example, the following commands load two traces into memory for
bank 1 of a module in slot 1. In this example, each byte of the LWORd to
output is sent as a separate byte.
TRAC:DATA:DIG:LWOR (@1101), MyTrace1, 255, 200, 128, 0
TRAC:DATA:DIG:LWOR (@1101), MyTrace2, 254, 192, 64, 32
You can also send trace data in IEEE- 488 block format using this
command.
The 34950A also has two special built- in traces for your use. You can
generate and download a count- up trace and a walking 1’s pattern using
the
TRACe:DATA:DIGital:FUNCtion
command. See the
Programmer's
Reference Help file
for more details.
Set which trace to use.
The
SOURce:DIGital:MEMory:TRACe
command
assigns the desired trace to the bank. This command allows you to switch
between the traces pre- loaded into the bank’s memory.
Enable the memory.
Enable the memory on the bank using the
SOURce:DIGital:MEMory:ENABle
command. This command sets the
selected trace to be the output and puts the bank in the wait- for-
trigger state.
N O T E
You can generate a count-down or walking zero pattern by inverting the
data line polarity.
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...