Features and Functions
2
34980A User’s Guide
55
•
For scanning, the specified sample count sets the number of readings
per channel (same for all channels in the scan list). If no channels have
been assigned to the scan list, the sample count sets the number of
readings per trigger for the internal DMM.
•
You can specify a sample count in conjunction with a trigger count and
a sweep count. The three parameters operate independent of one
another, and the total number of readings returned will be the product
of the three parameters.
•
You can store at least 500,000 readings in memory and all readings are
automatically time stamped. If memory overflows, the new readings will
overwrite the first (oldest) readings stored; the most recent readings are
always preserved.
•
The
CONFigure
and
MEASure?
commands automatically set the sample
count to 1.
•
The instrument sets the sample count to 1 after a Factory Reset
(
*RST
command). An Instrument Preset (
SYSTem:PRESet
command) or
Card Reset (
SYSTem:CPON
command) does
not
change the setting.
Front Panel Operation:
Scan (Configure) > SAMPLE COUNT
Remote Interface Operation:
SAMPle:COUNt
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...