68
34980A User’s Guide
2
Features and Functions
Alarm Limits
The instrument has four alarms which you can configure to alert you
when a reading exceeds specified limits on a channel during a scan.
You can assign a high limit, a low limit, or both to any configured channel
in the scan list. You can assign multiple channels to any of the four
available alarms (numbered 1 through 4). For example, you can configure
the instrument to generate an alarm on the Alarm 1 output when a limit
is exceeded on any of channels 1003, 2025, or 3020.
You can also assign alarms to channels on the digital modules (34950A
and 34952A). For example, you can generate an alarm when a specific bit
pattern or bit pattern change is detected on a digital input channel or
when a specific count is reached on a totalizer channel. With the digital
modules, the channels
do not
have to be part of the scan list to generate
an alarm. For complete details, see
“Using Alarms With the Digital
Modules”
on page 76.
Alarm data can be stored in one of two locations depending on whether a
scan is running when the alarm occurs.
1
If an alarm event occurs on a channel as it is being scanned, then that
channel’s alarm status is stored in reading memory as the readings are
taken. Each reading that is outside the specified alarm limits is logged
in memory. You can store at least 500,000 readings in memory during
a scan. You can read the contents of reading memory at any time, even
during a scan. Reading memory is
not
cleared when you read it.
2
As alarm events are generated, they are also logged in an alarm queue,
which is separate from reading memory. This is the only place where
non- scanned alarms get logged (alarms during a monitor, alarms
generated by the digital modules, etc.). Up to 20 alarms can be logged
in the alarm queue. If more than 20 alarm events are generated, they
will be lost (only the first 20 alarms are saved). Even if the alarm
queue is full, the alarm status is still stored in reading memory during
a scan. The alarm queue is cleared by the
*CLS
(clear status) command,
when power is cycled, and by reading all of the entries. A Factory Reset
(
*RST
command)
does not
clear the alarm queue.
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...