Interrupt Lines
Each bank has an interrupt line that can be used with memory input or
output operations. When a bank is set to input data, the interrupt line is
an output. When a bank is set to output data, the interrupt line is set to
be an input. You can set the polarity of the interrupt line for input and
output operations using the
CONFigure:DIGital:INTerrupt:POLarity
command.
You can configure the interrupt line drive mode, output drive level, and
input threshold. These parameters are set for both the handshake lines
and interrupt line on a bank. See
page 271
for details about setting these
parameters.
Memory Output Operations
For memory output operations, the interrupt line is sensed and can be
used to start or stop memory output operations. This provides a hardware
means to control the data output.
The
SOURce:DIGital:INTerrupt:MODE
command sets how the bank will
behave when using memory output. The mode can be set to one of three
values:
•
STARt
: The memory output will begin on the rising edge of the
interrupt line.
•
STOP
: The memory output is halted on the rising edge of the
interrupt line.
•
GATE
: The interrupt line acts a a gate for the memory output. The bank
can output when the interrupt line is asserted, and will stop when the
interrupt line is de- asserted.
When you have set the polarity and mode, enable the interrupt using the
SOURce:DIGital:INTerrupt:ENABle
command.
Memory Input Operations
For memory input operations, the interrupt line is an output and is set on
a pattern match or when the memory has been filled. You can set the
interrupt line to be driven or open collector using the
SENSe:DIGital:HANDshake:DRIVe
command.
N O T E
The settings for drive mode, output drive level, and input threshold also
apply to the bank’s handshake lines.
Содержание Agilent 34980A
Страница 1: ...Agilent Technologies Agilent 34980A Multifunction Switch Measure Unit User s Guide ...
Страница 2: ......
Страница 6: ......
Страница 198: ...186 34980A User s Guide 5 Matrix Switch Modules ...
Страница 210: ...198 34980A User s Guide 6 General Purpose Switch Modules ...
Страница 216: ...204 34980A User s Guide 7 RF Multiplexer Switch Modules ...
Страница 314: ...302 34980A User s Guide 11 4 Channel Isolated D A Converter with Waveform Memory Module ...
Страница 322: ...310 34980A User s Guide 12 Multifunction Module with DIO D A and Totalizer ...
Страница 326: ...314 34980A User s Guide 13 Breadboard Module 34959A Breadboard Module Layout shown with cover removed ...
Страница 335: ...Breadboard Module 13 34980A User s Guide 323 Dimensions of Suggested maximum size Custom PC Board ...
Страница 336: ...324 34980A User s Guide 13 Breadboard Module PC Board Footprint of Suggested DB50 Connectors M F ...